ARDMS SPI Exam Prep: Ultrasound Physics Study Guide

Updated Provider: American Registry for Diagnostic Medical Sonography (ARDMS)By the PrepWaypoint editorial team
Review status

Exam facts on this page were checked against the official provider on November 26, 2026. Study notes and practice questions were written by our editorial team and have not yet been reviewed by a licensed clinician or pharmacist. Use them for practice, not clinical decisions. How we write and check content.

The Sonography Principles and Instrumentation (SPI) exam from ARDMS is the ultrasound physics exam every ARDMS credential requires. To earn RDMS, RDCS, RVT or RMSKS, you pass SPI and a specialty exam within five years of each other. Doppler concepts are the largest area. This guide covers the outline, six original questions and a study plan.

On this page
  1. SPI content outline
  2. Ultrasound physics practice questions
  3. Study plan
  4. Official sources

SPI at a glance

Provider
American Registry for Diagnostic Medical Sonography (ARDMS), part of Inteleos
Content outline
SPI Content Outline, version 24.1
Length
About 110 multiple-choice questions
Time
2 hours
Fee
$275
Role
Required for every ARDMS credential; pair with a specialty exam within 5 years

SPI content outline

Doppler and image optimization together are 60% of the exam. Expect questions on aliasing, Nyquist limit, angle correction, gain, TGC, focus and artifacts.

  • Perform Ultrasound Examinations23%
  • Manage Ultrasound Transducers7%
  • Optimize Sonographic Images26%
  • Apply Doppler Concepts34%
  • Clinical Safety and Quality Assurance10%

Share of the exam. Source: ARDMS SPI Content Outline v24.1.

Physics to master

  • Wave properties: frequency, wavelength, propagation speed (1,540 m/s in soft tissue).
  • Resolution: axial, lateral, temporal.
  • Doppler: the Doppler equation, PRF, aliasing and spectral analysis.
  • Artifacts: shadowing, enhancement, reverberation, mirror image.
  • Bioeffects: ALARA, thermal and mechanical indices.

Ultrasound physics practice questions

These original questions follow the ARDMS outline. They are not ARDMS items.

6 original questions written for this page. Blueprint version: ARDMS SPI Content Outline, version 24.1. Written by the PrepWaypoint editorial team; not yet reviewed by a licensed clinician. Practice scores are not predictions of official results.

1.Apply Doppler Concepts

Aliasing appears on a spectral Doppler tracing. Which change helps eliminate it?

Show answer and explanation

Answer: A. Aliasing occurs when the Doppler shift exceeds the Nyquist limit (half the PRF). Raise the PRF, lower the frequency, shift the baseline or use continuous-wave Doppler.

B: Lowering PRF worsens aliasing.

C: Greater depth lowers the maximum PRF and worsens aliasing.

D: Higher frequency increases the Doppler shift and worsens aliasing.

2.Apply Doppler Concepts

At which angle between the beam and blood flow is no Doppler shift measured?

Show answer and explanation

Answer: B. The Doppler shift depends on cos θ; cos 90° = 0, so no shift is detected.

A: At 0 degrees the shift is maximal.

C: At 60 degrees a shift is measured.

D: At 45 degrees a shift is measured.

3.Perform Ultrasound Examinations

What is the assumed propagation speed of sound in soft tissue?

Show answer and explanation

Answer: C. Ultrasound systems assume 1,540 m/s in soft tissue to calculate depth.

A: 330 m/s is the speed in air.

B: 4,080 m/s is in the range of bone.

D: 1,000 m/s is not the standard value.

4.Optimize Sonographic Images

Which control adjusts brightness at specific depths to compensate for attenuation?

Show answer and explanation

Answer: D. TGC compensates for attenuation by adjusting receiver gain by depth.

A: Overall gain changes brightness at all depths equally.

B: Dynamic range changes contrast, not depth-specific brightness.

C: Output power changes transmitted energy and patient exposure.

5.Optimize Sonographic Images

Which artifact appears as bright echoes deep to a fluid-filled structure such as the gallbladder?

Show answer and explanation

Answer: A. Fluid attenuates less than surrounding tissue, so tissue behind it appears brighter: posterior enhancement.

B: Shadowing is a dark area behind highly attenuating structures.

C: Mirror image duplicates a structure across a strong reflector.

D: Side lobes place echoes off the main beam axis.

6.Clinical Safety and Quality Assurance

What does the ALARA principle mean in sonography?

Show answer and explanation

Answer: B. ALARA means minimizing output and dwell time while obtaining needed diagnostic information.

A: Higher output increases bioeffect risk.

C: Ultrasound is used in pregnancy when indicated.

D: Longer scans increase exposure.

Study plan

  1. Week 1: Baseline and wave basicsUnits, relationships and formulas.
  2. Weeks 2–3: DopplerEquation, aliasing fixes, color and spectral settings.
  3. Week 4: Image optimization and artifactsKnobs and their effects.
  4. Week 5: Transducers, safety and QAPhantoms, indices, care.
  5. Week 6: Full practiceA 2-hour timed session.

Other imaging and technical exams: NHA CET for EKG and CST for surgical technology.